Clean steam qenerator
A safer clean steam feeder
Coupled with the advance of medical care technologies and instruments, there is an evident need for improved steam purity in hospitals and medical care and other related industries. As an increasingly safe high quality steam is required, Miura has devoted its expertise to the development of a steam feeder equipment "Reboiler RB-S" that takes into account sterilization and humidification needs.
Clean steam qenerator

Example of clean steam feeder system
Miura, the Japanese top boiler manufacturer, proposes clean and safe steam system suitable for medical institutions using its boiler technology

RB-S Reboiler Features
  • Treated according to the feed water (pure water or RO water), a clean steam free of reagents (carcinogens) or rust is fed into a high pressure steam sterilizer
  • As related equipment can be installed close to the high pressure steam sterilizer, low cost clean steam sterilization using stainless pipes is possible
  • The total clean steam feed system, composed of reboiler, boiler, water treatment plant, and high pressure steam sterilizer, has the overall Miura support

Example of sterilizing steam composition analysis in a hospital

Item
Unit
EN285
Steam analysis
example *
Steam analysis
example **
Clean steam analysis
example for
SR-V type
Silica(SiO2) mg/L <=0.1 <0.1 <0.1 <0.1
Iron mg/L <=0.1 1.8 0.03 <0.03
Cadmium mg/L <=0.005 <0.005 <0.005 <0.005
Lead mg/L <=0.05 <0.1 <0.1 <0.1

Precious metals
(Zinc)

mg/L <=0.1 2.5 <0.05 0.02
Chloride ion mg/L <=0.1 0.2 <0.1 <0.1
Phosphorous mg/L <=0.1 <0.05 <0.05 <0.05
Electro conductivity µS/cm <=3 13.4 168 1.54
ph - 5-7 5.6 9.2 5.9

NOTE: * **Steam analysis example refers to common steam
(Japanese Association for Operative Medicine 21st Report)